GUI reference
R&S
®
ZNA
901User Manual 1178.6462.02 ─ 29
Physical
interface
(connector)
Interface
(protocol)
Address Remarks
USB USB-VISA
<ManID>::<ProdID>::<SerialNo>
e.g. 0x0AAD::0x0047::100098
2733 (0x0AAD) is the manufacturer ID of
Rohde & Schwarz.
71 (0x0047) is an example for a R&S prod-
uct ID (R&S SMF100A).
The serial number is device-specific.
LAN or USB Other Interface-specific, e.g. for SOCKET:
TCPIP0::<IpAddress>::<PortNo>::SOCKET
For USB-VISA:
USB0::<ManID>::<ProdID>::<SerialNo>::INSTR
Use complete VISA resource string.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
5.19.6 Frequency Converter tab
Allows you to set up the R&S ZNA for measurements using external frequency con-
verters.
The buttons on the "Frequency Converter" tab open the following dialogs
"Frequency Converter..." opens a dialog that allows you to define the "Converter
Topology", i.e. which converters are used an how they are connected to the
R&S ZNA. See Chapter 5.19.6.1, "Converter Configuration dialog", on page 902.
"Default Cable and Splitter Losses ..." opens a dialog that lets you specify the los-
ses between the IF and LO ports of the VNA and the converter. See Chap-
ter 5.19.6.3, "Default Cable and Splitter Losses dialog", on page 906.
"Save Converter Topology..."/"Load Converter Topology..." open standard file save/
open dialogs that allow you to save the current converter topology to a file or to
load a previously defined converter topology from a file (*.zcxml).
"Leveling Dataset" opens a dialog that allows you to provide leveling information to
the configuration. See Chapter 5.19.6.4, "Leveling Datasets dialog", on page 907.