Command reference
R&S
®
ZNA
1662User Manual 1178.6462.02 ─ 29
AHIGh | ALOW
High (low) voltage level in active state and low (high) level other-
wise
Note that pulse generator pulses cannot be indicated using high
and low voltage levels.
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
on page 1660.
Options: R&S ZNA-B91
Manual operation: See "Signal Type" on page 564
7.3.19.3
Other commands
TRIGger:STATus:READy? [<TrigSource>]
Queries the current "Ready for Trigger" state
Query parameters:
<TrigSource> ANY
For Keysight PNA Compatibility, ANY (and only ANY) is accep-
ted as parameter.
Usage: Query only
TRIGger:WAIT:READy? [<Timeout>]
Returns 1 when the R&S ZNA is "Ready for Trigger", or 0 after the specified <Time-
out>.
Query parameters:
<Timeout> Timeout value (optional, default 1 ms)
Default unit: s
Usage: Query only
7.3.20 New commands for R&S ZNA
● Mixer phase and delay measurements............................................................... 1663
● Advanced port settings........................................................................................1668
● Internal 2nd LO (R&S ZNA-B5)...........................................................................1670
● Other................................................................................................................... 1670
SCPI command reference