Slow down the sweep: select a smaller bandwidth or increase the "Sweep Time"
setting.
Suffix:
<Ch>
.
Channel number
<n> 1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<TrigOutDuration> Pulse width between 10 ns (1E-8 s) and 30 s.
*RST: 1 us
Default unit: s
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
on page 1661
Options: R&S ZNA-B91
Manual operation: See "Width" on page 564
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
<Type>
Selects the sweep interval to be indicated at the related Trigger Out connector.
Suffix:
<Ch>
.
Channel number
<n> 1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<Type> NONE | SWEep | SEGMent | PPOint | POINt | ACQuisition |
PGENerator
NONE
No trigger out signal for this channel and Trigger Out connector.
SWEep | SEGMent | PPOint | POINt
A trigger out signal is generated for every sweep, segment, par-
tial measurement or sweep point, respectively.
ACQuisition
A trigger out signal is generated for every data acquisition phase
PGENerator
For pulse generator pulses, pulses with identical or opposite
polarity can be generated as trigger out signal
*RST:
SWEep
Example:
TRIGger:CHANnel1:AUXiliary1:STYPe POSitive
Generate pulses with positive polarity at Trigger Out A.
TRIGger:CHANnel:AUXiliary:INTerval SWEep
Send one output trigger pulse per sweep.
TRIGger:CHANnel:AUXiliary:POSition BEFore
TRIGger:CHANnel1:AUXiliary1:DELay 1 us
TRIGger:CHANnel:AUXiliary:DURation 10 us
Send the output trigger 1 μs before sweep start, and with 10 μs
pulse width.