Command reference
R&S
®
ZNA
1661User Manual 1178.6462.02 ─ 29
Options:  R&S ZNA-B91
Manual operation:  See "Sequence" on page 564
TRIGger:CHANnel<Ch>:AUXiliary<n>:POSition <Type>
If pulses are used as trigger out signals (see TRIGger:CHANnel<Ch>:
AUXiliary<n>:STYPe), the pulses can either indicate the "Start" or "End" of the rela-
ted sweep interval (see TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval).
Suffix: 
<Ch>
.
Channel number
<n> 1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<Type> BEFore | AFTer
BEFore
is appropriate if the device needs to be triggered before the data
is acquired (typical example: a power meter which is used for
power calibration)
AFTer
is appropriate if the device needs to be triggered just after data
acquisition (typical example: an external generator which must
be reconfigured to get ready for the next measurement)
*RST:  AFTer
Example:  See TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
on page 1661
Options:  R&S ZNA-B91
Manual operation:  See "Position" on page 565
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe <Type>
Defines how the selected sweep sequence is indicated at the related Trigger Out con-
nector.
Use TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval to select the sweep
sequence.
Suffix: 
<Ch>
.
Channel number
<n> 1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<Type> POSitive | NEGative | AHIGh | ALOW
POSitive | NEGative
pulses with positive or negative polarity
SCPI command reference