Messages in the dialog
An information message (or error message) is displayed if one of the following hap-
pens:
●
One of the selected calibration kits is described by ideal kit parameters or typical
values.
●
One of the selected calibration kits does not contain all standards that are required
for the chosen calibration type.
●
Different connector types are defined at the ports but the selected calibration type
requires uniform connectors.
●
A cal kit standard, cal unit or power meter does not cover the entire calibrated fre-
quency range.
●
A combination of cal units and cal kits is configured in a channel
Calibration Type tab
The "Calibration Type" tab allows you to configure the calibration setup (i.e. a set of
calibration definitions) for the channel selected in the title area of the calibration setup
dialog.