GUI reference
R&S
®
ZNA
584User Manual 1178.6462.02 ─ 29
Use Power Cal
If checked, you can select the power calibrations to be performed together with the
system error correction.
The default setting and the selected power cals depend on the active channel mode.
E.g., for a standard (S-parameter) channel a SMARTerCal is proposed if a power
meter is available.
Source Leveling/Src. Lev. Method
Adds a Source power calibration of the related port to the current calibration definition,
which comes in handy for SMARTer calibrations (see Chapter 4.5.7.3, "Combining
SMARTerCal with scalar power calibration", on page 226).
When you run the Calibration wizard, an additional source flatness calibration step is
performed.
"Src. Lev. Method" is equivalent to "Pwr Cal Method" on the "Power Cal Settings" tab
(see "Flatness Cal – Pwr Cal Method" on page 628).
Remote command:
SOURce:POWer:CORRection:COLLect:METHod
Connector / Gender
Defines the connector types and genders of the ports to be calibrated. For symmetric
(sexless) connector types (e.g. 7 mm / PC7), "Gender" is unavailable.
If "Default Conn." is active, the "Default Connector" is set in the "Default for all Chan-
nels" section (to the left of the Cal. Settings table). If "Same Gen." is active, the gen-
ders at all ports are always adjusted to the current selection.
User-defined connectors can be added or removed in the Cal Connector Types dialog.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>
[SENSe<Ch>:]CORRection:CONNection
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe<Ch>:]CORRection:CONNection:DELete
Add
Use "Add" to add a calibration definition to the selected channel. The default calibration
type depends on the channel's measurement mode.
Delete
Select a row in the Cal. Settings table and "Delete" the corresponding calibration defi-
nition.
Cal. Settings
The "Cal. Settings" table displays the calibration definitions of the current channel. The
available columns depend on the available calibration devices.
Cal softtool