GUI reference
R&S
®
ZNA
585User Manual 1178.6462.02 ─ 29
Figure 5-24: Cal settings for standard calibration units
Figure 5-25: Cal settings for inline calibration systems
Note: Currently, for each channel either manual calibrations, automatic calibrations, or
inline calibration systems are allowed.
Port <i> ← Cal. Settings
Tick the checkboxes in the "Port <i>" columns to select the ports to be calibrated in the
corresponding calibration.
Note: Calibration and port de-/activation.
The analyzer fimware automatically activates/deactivates ports during/after a (success-
ful) calibration:
●
Calibrated ports that were previously disabled, are automatically enabled as single-
ended logical ports.
●
An uncalibrated port that is not used by a measurement (i.e. the port is not required
by any trace of the related channel) is disabled.
Source Port ← Cal. Settings
Source port for unidirectional n-port calibrations such as "One Path Two Ports" or
"Trans Norm".
Remote command:
First test port in various autocal definition commands such as [SENSe<Ch>:
]CORRection:COLLect:AUTO:TYPE
Common Port ← Cal. Settings
This column is only available if "Use Reduced Number of Through" on page 859 is
activated in the "Calibration" tab of the "System Config" dialog.
For full n-port automatic calibrations, it defines the port that must appear in each port
assignment, and that are as the "center" of all Through measurements.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Calibration Type ← Cal. Settings
Displays the calibration type.
Tap the table cell to open the "Define Calibration" dialog that lets you choose calibrated
ports and type of the calibration definition. The available calibration types depend on
the channel mode of the active channel.
Cal softtool