GUI reference
R&S
®
ZNA
586User Manual 1178.6462.02 ─ 29
Figure 5-26: Define Calibration dialog: manual calibration
Figure 5-27: Define Calibration dialog: automatic calibration
The green arrow symbols give a preview of the type and the number of calibration
sweeps involved:
Curved arrows (example: "Refl Norm Open") denote one or more reflection mea-
surements at each port.
Reflection calibration types can be used for any set of test ports: reflection calibra-
tions are repeated for each port.
Straight, horizontal arrows (example: "Trans Norm") denote one or more transmis-
sion measurements between each pair of two ports.
Transmission calibration types require at least two physical ports. For unidirectional
transmission calibration types ("Trans Norm", "One Path Two Ports"), the
("Source" port) must be specified in addition.
The full n-port calibration types (n > 1, e.g. "TOSM") are symbolized by a closed
square symbol. The number of arrows increases the complexity but can also
improve the accuracy of the calibration.
Cal softtool