GUI reference
R&S
®
ZNA
587User Manual 1178.6462.02 ─ 29
For an overview, refer to Table 4-8.
Remote command:
Manual calibration
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
Automatic calibration
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Power Meter ← Cal. Settings
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See "Configured Devices" on page 893.
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Cal Kit Port <i> ← Cal. Settings
For each physical port covered by the selected calibration definition, you can select a
cal kit, cal unit, or inline calibration system. The drop-down list contains the kits for the
selected connector type and the connected calibration units.
The assignment of a calibration kit or unit to a connector type must be the same for all
physical ports covered by the calibration definition. If a calibration kit is changed, the
R&S ZNA automatically assigns the new kit to all ports with the same connector type.
Use Import Cal Kit... to add new kits to the list.
Remote command:
[SENSe:]CORRection:CKIT:SELect
Characterization ← Cal. Settings
Displays all characterizations that are stored in the active cal unit. The "Factory" char-
acterization is available for all calibration units; it ensures an accurate calibration for all
standard applications. To account for modifications of the cal unit such as the connec-
tion of additional adapters, you can generate modified sets of characterization data
using the cal unit characterization wizard. See Chapter 5.11.2.3, "Characterize Cal Unit
dialog", on page 621. By default, the R&S ZNA uses the last generated cal unit char-
acterization.
For an inline calibration system, each connected inline calibration unit has its own set
of characterizations:
Cal softtool