Command reference
R&S
®
ZNA
1392User Manual 1178.6462.02 ─ 29
Example: See Chapter 8.2, "Condensed programming examples",
on page 1731.
Usage: Event
Options: R&S ZNA-K30
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt <arg0>, <arg1>, <arg2>,
<arg3>, <arg4>
Initiates a noise figure calibration for the related channel.
The calibration sweeps are started using [SENSe<Ch>:]CORRection:COLLect:
NFIGure[:ACQuire], the noise figure calibration is terminated using [SENSe<Ch>:
]CORRection:COLLect:NFIGure:SAVE or [SENSe<Ch>:]CORRection:
COLLect:NFIGure:END.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
The subsequent noise figure calibration commands must use the
same channel number.
Setting parameters:
<arg0> Number of the physical VNA port connected to the DUT output.
<arg1> Number of the physical VNA port connected to the DUT input.
<arg2> ON | OFF - Boolean parameter, indicating whether an external
attenuator is used.
<arg3> Attenuation of the source level during the source noise calibra-
tion.
<arg4> Attenuation of the source level during the measurement (i.e.
after calibration).
Example: See Chapter 8.2, "Condensed programming examples",
on page 1731.
Usage: Setting only
Options: R&S ZNA-K30
[SENSe:]CORRection:COLLect:PMETer:ID <PowerMeter>, <TestPort>
Selects an external power meter for the SMARTerCal and assigns it to an analyzer
port.
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
SCPI command reference