Command reference
R&S
®
ZNA
1391User Manual 1178.6462.02 ─ 29
RECeiver – a receiver noise calibration is performed
SOURce – a source noise calibration is performed
ATTenuator – noise calibration of the external attenuator is per-
formed (optional, only if an external attenuator is used; see
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt).
The calibration steps have to be performed in the order SOURce
→ RECeiver → ATTenuator.
Example: See Chapter 8.2, "Condensed programming examples",
on page 1731.
Usage: Setting only
Options: R&S ZNA-K30
[SENSe<Ch>:]CORRection:COLLect:NFIGure:END
Terminates a noise figure calibration, discarding the acquired calibration data.
A noise figure calibration is initiated using [SENSe<Ch>:]CORRection:COLLect:
NFIGure:STARt and data are aquired using [SENSe<Ch>:]CORRection:
COLLect:NFIGure[:ACQuire]. It must be terminated by one of the commands
[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE or
[SENSe<Ch>:]CORRection:COLLect:NFIGure:END.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
All commands related to a noise figure calibration must use the
same channel number.
Example: See Chapter 8.2, "Condensed programming examples",
on page 1731.
Usage: Event
Options: R&S ZNA-K30
[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE
Completes a noise figure calibration, storing and applying the acquired calibration data.
A noise figure calibration is initiated using [SENSe<Ch>:]CORRection:COLLect:
NFIGure:STARt and data are aquired using [SENSe<Ch>:]CORRection:
COLLect:NFIGure[:ACQuire]. It must be terminated by one of the commands
[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE or [SENSe<Ch>:
]CORRection:COLLect:NFIGure:END.
In case a noise measurement is missing, an error is reported. In this case, simply per-
form the missing ACQ step, then repeat SAVE.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
All commands related to a noise figure calibration must use the
same channel number.
SCPI command reference