Command reference
R&S
®
ZNA
1390User Manual 1178.6462.02 ─ 29
Parameter Type (GUI)
VUOSm UOSM Vector Mixer
TRL | TOM | TSM | TRM | TNA TRL | TOM | TSM | TRM | TNA
ARPTosm | POPTport | PTOSm |
PUOSm | VPUosm | PTRL | PTOM |
PTSM | PTRM | PTNA
One of the previous n-port calibration types, in combination with a
receiver power calibration at test port <TestPort1> (SMARTerCal)
PFTRans, PRTRans, PFRTrans FTRans, RTRans, in combination with a receiver power calibration
at the source port(s)
PREFl, PRSHort, PFOPort REFl, RSHort, FOPort, in combination with a receiver power cali-
bration at all ports
PRRCv Reference receiver power calibration at all ports
CPUOsm PUOSM Scalar Vector Mixer
MTOSm | MUOSm | MOSM | MTRL METAS Calibration of type OSM, UOSM, OSM, or TRL
MDELay Mixer Delay
Delay mixer calibration (stand-alone) for two-tone group delay
measurements
UMDelay UOSM Mixer Delay calibration
UOSM combined with delay mixter calibration for two-tone group
delay measurements (R&S ZNA-K9)
PUMDelay UMDelay calibration with complementary source flatness cal for
two-tone group delay measurements (R&S ZNA-K9)
[SENSe<Ch>:]CORRection:COLLect:NFIGure[:ACQuire] <arg0>
Starts one of the three noise calibration sweeps that are required for a noise figure cali-
bration of the related channel.
The noise figure calibration is initiated using [SENSe<Ch>:]CORRection:COLLect:
NFIGure:STARt and must be terminated by one of the commands [SENSe<Ch>:
]CORRection:COLLect:NFIGure:END or [SENSe<Ch>:]CORRection:
COLLect:NFIGure:SAVE.
Note: A noise figure calibration must be terminated with one of the following com-
mands, to be sent immediately after the sequence of ...[:ACQuire] commands:
Abort the noise figure calibration, discarding the acquired calibration data.
Terminate the noise figure calibration, applying the acquired calibration data (the
active channel is calibrated) These commands ensure that the analyzer is able to
start a new measurement sweep.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
The other noise figure calibration commands must use the same
channel number.
Setting parameters:
<arg0> RECeiver | SOURce | ATTenuator