GUI reference
R&S
®
ZNA
543User Manual 1178.6462.02 ─ 29
Auto | ... | Slow
The R&S ZNA firmware implements three algorithms for calculating hardware settling
wait times (see "Wait time control algorithms" on page 115). Choose one of these algo-
rithms, or let the R&S ZNA decide which one to use ("Auto" = default).
Calculated wait times are point-specific and not constant. In controlled timing mode,
however, this is a must. To achieve constant wait times, the firmware chooses the max-
imum wait time calculated for the measurement points as the constant wait time.
In manual mode mode, you can overwrite or limit the calculated values.
Remote command:
[SENSe<Ch>:]SWEep:TIME:MODE
Manual
"Active" allows you to activate or deactivate manual wait time control, "Settings..."
opens a dialog that allows you to define manual wait times (see Chapter 5.10.2.2,
"Manual Wait Time Settings dialog", on page 543).
Remote command:
[SENSe<Ch>:]SWEep:CTIMing:MODE
Show Sweep Timing
Show/hides the sweep timing info box.
See Table 4-2.
Note: The firmware does not display the sweep timing info box, if the display of instru-
ment messages is blocked (see "Show Instrument Messages" on page 864).
Remote command:
[SENSe<Ch>:]SWEep:CTIMing:VISualize
5.10.2.2 Manual Wait Time Settings dialog
Allows you to activate and configure manual wait times for different channel/sweep
events. By selectively activating manual values for certain events, you can also config-
ure a mixture of calculated and manual wait times.
Sweep Softtool