GUI reference
R&S
®
ZNA
544User Manual 1178.6462.02 ─ 29
In controlled timing mode, the firmware chooses the maximum total wait time calcula-
ted or set for the measurement points as the constant wait time. For the first point in a
channel, the total wait time comprises the calculated wait time for channel initialization
(Channel Begin), which cannot be set in controlled timing mode.
Access: Channel – [Sweep] > "Wait Time Control" tab > "Manual Settings" section >
"Settings..."
Active
Activates or deactivates manual wait time control for the current channel. Same func-
tionality as the "Active" checkbox on theWait Time Control tab.
If checked, you can activate and configure manual wait times for various channel and
sweep events.
Remote command:
[SENSe<Ch>:]SWEep:CTIMing:MODE
Channel Begin
Allows you to activate and specify manual wait times for the begin of the current chan-
nel. A distinction is made between:
"same" – the current channel is the first one to be swept, or the same as the previ-
ously swept one
"other" – the current channel is different from the previously swept one
Manual wait times for "same" and "other" can be activated separately, and different
minimum wait times can be specified. The (common) lower limit for the minimum wait
times is determined by certain HW properties and enforced by the analyzer firmware.
Sweep Softtool