GUI reference
R&S
®
ZNA
566User Manual 1178.6462.02 ─ 29
●
In "Single" sweep mode, the measurement is stopped after the configured number
of sweeps.
Restart Sweep initiates a new measurement cycle.
For the DEFAULT Remote Language, this setting applies to the active channel only.
Use All Channels Continuous/All Channels on Hold to select a common sweep mode
for all channels.
For remote languages other than DEFAULT, this setting applies to all channels.
Remote command:
INITiate<Ch>:CONTinuous
See also:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
CONFigure:CHANnel:MEASure:ALL[:STATe]
Controlled Timing
Activates/deactivates the controlled timing mode, which is provided by Continuous data
recording option R&S ZNA-K28.
Note: This checkbox is only available if option R&S ZNA-K28 is installed. It is only
enabled if the current setup contains a single channel.
You can control the wait times from the Wait Time Control tab.
Remote command:
INITiate<Ch>:CTIMing
Sweeps/Memory Size
"Sweeps" defines the number of sweeps to be performed (and buffered) in "Single"
sweep mode (see "Continuous / Single" on page 565). In "Continuous" sweep mode,
and if Controlled Timing is inactive, it is not available (grayed out).
In "Continuous" sweep mode, if sweep control is active, the label changes to "Memory
Size", indicating how many sweeps are buffered (ring buffer).
For remote languages other than DEFAULT, the "Sweeps" setting is disabled. You can
define the number of sweeps in the Restart Manager dialog.
Remote command:
[SENSe<Ch>:]SWEep:COUNt
[SENSe:]SWEep:COUNt:ALL
Restart Sweep
Stops the current measurement cycle and starts a new one.
For the DEFAULT Remote Language, the effect of this command depends on the
sweep modes of the active channel, and the restart behavior defined in the Restart
Manager dialog:
Sweep Softtool