Command reference
R&S
®
ZNA
1502User Manual 1178.6462.02 ─ 29
Parameters:
<Sweep> Number of consecutive sweeps to be measured and buffered.
Range: 1 to 100000
*RST: 1
Example: See CALCulate<Chn>:DATA:NSWeep:FIRSt?
Manual operation: See "Sweeps/Memory Size" on page 566
[SENSe:]SWEep:COUNt:ALL <Sweep>
Defines the number of sweeps to be measured in single sweep mode
(INITiate<Ch>:CONTinuous OFF). The setting is applied to all channels. Use
[SENSe<Ch>:]SWEep:COUNt to define the sweep count for a single channel.
Setting parameters:
<Sweep> Number of consecutive sweeps to be measured.
Range: 1 to 100000
*RST: 1
Example: See Chapter 8.2.4, "Data handling", on page 1743.
Usage: Setting only
Manual operation: See "Sweeps/Memory Size" on page 566
[SENSe<Ch>:]SWEep:CTIMing:CHANnel <MinWaitTime>
Sets/gets the minimum wait time for the initialization of channel <Ch>, in case it is the
first channel to be swept, or the same as the previously swept channel.
Only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:CHANnel:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
In controlled timing mode, this setting is ignored.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> <MinWaitTime>
Minimum wait time
Default unit: s
Manual operation: See "Channel Begin" on page 544
[SENSe<Ch>:]SWEep:CTIMing:CHANnel:MODE <Boolean>
Activates or deactivates manual wait times for the initialization of channel <Ch>, in
case it is the first channel to be swept, or the same as the previously swept channel.
SCPI command reference