Command reference
R&S
®
ZNA
1503User Manual 1178.6462.02 ─ 29
The minimum wait time can be set using [SENSe<Ch>:]SWEep:CTIMing:CHANnel.
It only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:CHANnel:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE on page 1504 are set to ON.
In controlled timing mode, this setting is ignored.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "Channel Begin" on page 544
[SENSe<Ch>:]SWEep:CTIMing:COMPression <WaitTime>
Sets/gets the wait time for the first point of each 2D gain compression (power) sub-
sweep in channel <Ch> (see [SENSe<Ch>:]FREQuency:COMPression:POINt).
Only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:COMPression:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> <WaitTime>
Wait time (fixed)
Default unit: s
Manual operation: See "First Point of Segments" on page 545
[SENSe<Ch>:]SWEep:CTIMing:COMPression:MODE <Boolean>
Activates or deactivates manual wait times for the first point of each 2D gain compres-
sion (power) subsweep in channel <Ch> (see [SENSe<Ch>:]FREQuency:
COMPression:POINt).
The wait time can be set using [SENSe<Ch>:]SWEep:CTIMing:COMPression. It
only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:COMPression:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE on page 1504 are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "First Point of Segments" on page 545
SCPI command reference