1504User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]SWEep:CTIMing:MODE <Boolean>
Activates or deactivates manual wait time settings for channel <Ch>.
Note that the manual wait times for the related channel/sweep events must activated
separately using
[SENSe<Ch>:]SWEep:CTIMing:CHANnel:MODE
[SENSe<Ch>:]SWEep:CTIMing:OCHannel:MODE
[SENSe<Ch>:]SWEep:CTIMing:SEGMent:MODE
[SENSe<Ch>:]SWEep:CTIMing:COMPression:MODE
[SENSe<Ch>:]SWEep:CTIMing:PORT:MAX:MODE
[SENSe<Ch>:]SWEep:CTIMing:PORT:MIN:MODE
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "Manual" on page 543
[SENSe<Ch>:]SWEep:CTIMing:OCHannel <MinWaitTime>
Sets/gets the minimum wait time for the initialization of channel <Ch>, in case it is dif-
ferent from the previously swept one.
Only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:OCHannel:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
In controlled timing mode, this setting is ignored.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> <MinWaitTime>
Minimum wait time
Default unit: s
Manual operation: See "Channel Begin" on page 544
[SENSe<Ch>:]SWEep:CTIMing:OCHannel:MODE <Boolean>
Activates or deactivates manual wait times for the initialization of channel <Ch>, in
case it is different from the previously swept one.
The minimum wait time can be set using [SENSe<Ch>:]SWEep:CTIMing:OCHannel
. It only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:OCHannel:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE on page 1504 are set to ON.
SCPI command reference

Table of Contents