Command reference
R&S
®
ZNA
1506User Manual 1178.6462.02 ─ 29
Only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:PORT:MIN:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> <WaitTime>
Minimum wait time
Default unit: s
Manual operation: See "First Point of Ch / Drive Port Change" on page 545
[SENSe<Ch>:]SWEep:CTIMing:PORT:MIN:MODE <Boolean>
Activates or deactivates manual minimum wait times for the first point of channel <Ch>,
and the first point after a drive port change in channel <Ch>.
The minimum wait time can be set using [SENSe<Ch>:]SWEep:CTIMing:PORT:MIN
. It only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:PORT:MIN:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "First Point of Ch / Drive Port Change" on page 545
[SENSe<Ch>:]SWEep:CTIMing:SEGMent <WaitTime>
Sets/gets the wait time for the first point of each segmented frequency sweep in chan-
nel <Ch> ([SENSe<Ch>:]SWEep:TYPE SEGMent).
Only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:SEGMent:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> <WaitTime>
Wait time (fixed)
Default unit: s
Manual operation: See "First Point of Segments" on page 545