Command reference
R&S
®
ZNA
1507User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]SWEep:CTIMing:SEGMent:MODE <Boolean>
Activates or deactivates manual wait times for the first point of each segmented fre-
quency sweep in channel <Ch> ([SENSe<Ch>:]SWEep:TYPE SEGMent).
The wait time can be set using [SENSe<Ch>:]SWEep:CTIMing:SEGMent. It only
takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:SEGMent:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE on page 1504 are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "First Point of Segments" on page 545
[SENSe<Ch>:]SWEep:CTIMing:STIMe
<Value>
Sets/gets a user-defined wait time per measurement point in channel <Ch>.
Only takes effect, if [SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE is set to USER
and [SENSe<Ch>:]SWEep:CTIMing:MODE is set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<Value> Wait time (fixed)
*RST: 1 ms
Default unit: s
Manual operation: See "All Points" on page 545
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE <Mode>
Enables/disables user-defined wait times per measurement point in channel <Ch>.
Note
●
Setting <Mode> to USER or LIMited automatically activates manual wait time con-
trol for channel <Ch> ([SENSe<Ch>:]SWEep:CTIMing:MODE ON).
●
Deactivating manual wait time control for channel <Ch> ([SENSe<Ch>:]SWEep:
CTIMing:MODE OFF) disables all user-defined wait times in this channel, whether
active or not.
Suffix:
<Ch>
.
Channel number
Parameters:
<Mode> AUTO | USER | LIMited
SCPI command reference