GUI reference
R&S
®
ZNA
545User Manual 1178.6462.02 ─ 29
This setting is not available in controlled timing mode.
Remote command:
"same":
[SENSe<Ch>:]SWEep:CTIMing:CHANnel:MODE
[SENSe<Ch>:]SWEep:CTIMing:CHANnel
"other":
[SENSe<Ch>:]SWEep:CTIMing:OCHannel:MODE
[SENSe<Ch>:]SWEep:CTIMing:OCHannel
First Point of Ch / Drive Port Change
Activates or deactivates manual wait times for the current channel's first point, and its
points following a drive port change. Minimum and maximum wait times can be activa-
ted and specified separately. They are valid for both channel event types.
Remote command:
Minimum wait time:
[SENSe<Ch>:]SWEep:CTIMing:PORT:MIN:MODE
[SENSe<Ch>:]SWEep:CTIMing:PORT:MIN
Maximum wait time:
[SENSe<Ch>:]SWEep:CTIMing:PORT:MAX:MODE
[SENSe<Ch>:]SWEep:CTIMing:PORT:MAX
First Point of Segments
Separately activates and specifies exact manual wait times for:
●
The first point of a segment, if the current channel performs a segmented fre-
quency sweep
●
The first point of each (power) subsweep, if the current channel performs a two-
dimensional gain compression sweep
Only applies to those points that are not First Point of Ch / Drive Port Change.
Remote command:
Segmented frequency sweep:
[SENSe<Ch>:]SWEep:CTIMing:SEGMent:MODE
[SENSe<Ch>:]SWEep:CTIMing:SEGMent
2-dimensional gain compression measurements:
[SENSe<Ch>:]SWEep:CTIMing:COMPression:MODE
[SENSe<Ch>:]SWEep:CTIMing:COMPression
All Points
Activates and specifies manual wait times for other sweep points.
●
"Auto" deactivates manual settings
●
"User Def" allows you to specify an exact value
●
"Limited" allows you to specify an allowed range
Remote command:
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE
[SENSe<Ch>:]SWEep:CTIMing:STIMe
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MIN
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MAX
Sweep Softtool