Command reference
R&S
®
ZNA
1508User Manual 1178.6462.02 ─ 29
AUTO
The wait time is calculated automatically, with the algorithm
selected using [SENSe<Ch>:]SWEep:TIME:MODE.
USER
A user-defined wait time can be specified using [SENSe<Ch>:
]SWEep:CTIMing:STIMe.
LIMited
Upper and lower limits for the wait time can be specified using
[SENSe<Ch>:]SWEep:CTIMing:PORT:MAX:MODE and
[SENSe<Ch>:]SWEep:CTIMing:PORT:MIN:MODE.
Manual operation: See "All Points" on page 545
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MAX
<Value>
Sets/gets a the maximum wait time per measurement point in channel <Ch>.
Only takes effect, if [SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE is set to
LIMited and [SENSe<Ch>:]SWEep:CTIMing:MODE is set to ON.
The corresponding minimum can be set using [SENSe<Ch>:]SWEep:CTIMing:
STIMe:MIN.
Suffix:
<Ch>
.
Channel number
Parameters:
<Value> Wait time (fixed)
*RST: 1 ms
Default unit: s
Manual operation: See "All Points" on page 545
[SENSe<Ch>:]SWEep:CTIMing:STIMe:MIN <Value>
Sets/gets a the minimum wait time per measurement point in channel <Ch>.
Only takes effect, if [SENSe<Ch>:]SWEep:CTIMing:STIMe:MODE is set to
LIMited and [SENSe<Ch>:]SWEep:CTIMing:MODE is set to ON.
The corresponding maximum can be set using [SENSe<Ch>:]SWEep:CTIMing:
STIMe:MAX.
Suffix:
<Ch>
.
Channel number
Parameters:
<Value> Wait time (fixed)
*RST: 1 ms
Default unit: s
Manual operation: See "All Points" on page 545