Command reference
R&S
®
ZNA
1505User Manual 1178.6462.02 ─ 29
In controlled timing mode, this setting is ignored.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "Channel Begin" on page 544
[SENSe<Ch>:]SWEep:CTIMing:PORT:MAX
<WaitTime>
Sets/gets the maximum wait time for the first point of channel <Ch>, and the first point
after a drive port change in channel <Ch>.
Only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:PORT:MAX:MODE and
[SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> <WaitTime>
Maximum wait time
Default unit: s
Manual operation: See "First Point of Ch / Drive Port Change" on page 545
[SENSe<Ch>:]SWEep:CTIMing:PORT:MAX:MODE <Boolean>
Activates or deactivates manual maximum wait times for the first point of channel
<Ch>, and the first point after a drive port change in channel <Ch>.
The maximum wait time can be set using [SENSe<Ch>:]SWEep:CTIMing:PORT:
MAX. It only takes effect, if both [SENSe<Ch>:]SWEep:CTIMing:PORT:MAX:MODE
and [SENSe<Ch>:]SWEep:CTIMing:MODE are set to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "First Point of Ch / Drive Port Change" on page 545
[SENSe<Ch>:]SWEep:CTIMing:PORT:MIN <WaitTime>
Sets/gets the minimum wait time for the first point of channel <Ch>, and the first point
after a drive port change in channel <Ch>.
SCPI command reference