Command reference
R&S
®
ZNA
1509User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]SWEep:CTIMing:VISualize <Boolean>
Activates or deactivates the sweep timing info box for channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Manual operation: See "Show Sweep Timing" on page 543
[SENSe<Ch>:]SWEep:DETector:TIME
<DetectorTime>
Defines the observation time per sweep point if a detector other than the "Normal"
(SAMple) detector is used. The detector is selected together with the measured quan-
tity (CALCulate<Ch>:PARameter:MEASure or CALCulate<Ch>:PARameter:
SDEFine).
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime> Detector time
Range: 0 s to 3456000 s
*RST: 0.01 s
Default unit: s
Example:
*RST; :CALC:PAR:MEAS 'TRC1', 'a1D1AVG'
Select the wave quantity a
1
for the default trace and activate the
AVG detector.
SWE:DET:TIME 1
Specify an observation time of 1 s at each sweep point.
Manual operation: See "Detector" on page 350
[SENSe<Ch>:]SWEep:DWELl <MeasDelay>
Defines the "Meas. Delay" before each partial measurement or the first partial mea-
surement (see [SENSe<Ch>:]SWEep:DWELl:IPOint). Setting a delay disables the
automatic calculation of the (minimum) sweep time (see [SENSe<Ch>:]SWEep:
TIME:AUTO).
Suffix:
<Ch>
.
Channel number
SCPI command reference