Command reference
R&S
®
ZNA
1510User Manual 1178.6462.02 ─ 29
Parameters:
<MeasDelay> Measurement delay before each partial measurement. Changing
the delay leaves the number of points unchanged but has an
impact on the duration of the sweep ([SENSe<Ch>:]SWEep:
TIME).
Range: 0 s to 13680 s
*RST: 0 s
Default unit: s
Example:
*RST
Reset the instrument, activating a frequency sweep with the S-
parameter S
21
as a measurement result for channel and trace
no. 1.
SWEep:TIME?
Query total sweep time.
SWEep:DWEL 1
SWEep:DWEL:IPOint ALL
Set a delay of 1 s for each partial measurement.
SWE:TIME?
Query total sweep time. The time is extended by the delay times
the total number of sweep points (one partial measurement per
sweep point required).
Manual operation:
See "Meas Delay" on page 541
[SENSe<Ch>:]SWEep:DWELl:IPOint <InsertionPoints>
Defines whether the measurement delay (previously defined via [SENSe<Ch>:
]SWEep:DWELl) is inserted before all partial measurements or before the first partial
measurement only.
Suffix:
<Ch>
.
Channel number
Parameters:
<InsertionPoints> ALL | FIRSt
Insertion before all or before the first partial measurement
*RST: ALL
Example: See [SENSe<Ch>:]SWEep:DWELl
Manual operation: See "All Partial Meas'ments / First Partial Meas'ment"
on page 542
[SENSe<Ch>:]SWEep:FPOWer <Mode>
Selects the power sweep mode.
SCPI command reference