Command reference
R&S
®
ZNA
1266User Manual 1178.6462.02 ─ 29
Example: Alternative settings using the R&S ZVAB compatibility mode:
*RST; :SYSTem:LANGuage 'ZVABT'
*RST; :INIT:CONT OFF
Activate single sweep mode for all channels (including channel
no. 2 created later).
INIT:SCOP SING
State that a single sweep will be performed in the active chan-
nel.
CALC2:PAR:SDEF 'TRC2','S11'; :INIT2
Create channel no. 2 with a new trace and start a single sweep
in channel no. 2. Start a single sweep in the second channel.
Manual operation: See "Continuous / Single" on page 565
INITiate:CONTinuous:ALL <Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep
mode.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Parameters:
<Boolean> ON - the analyzer measures continuously in all channel, repeat-
ing the current sweep. The query returns ON (1) if at least one
channel is measured continuously.
OFF - the measurement is stopped after the number of sweeps
defined via [SENSe<Ch>:]SWEep:COUNt. INITiate<Ch>[:
IMMediate][:DUMMy] initiates a new measurement cycle.
*RST: ON
Example: See INITiate<Ch>:CONTinuous
Manual operation: See "All Channels Continuous/All Channels on Hold"
on page 567
INITiate<Ch>:CTIMing <Boolean>
Activates/deactivates the controlled timing mode.
Can only be executed if the current setup contains a single channel.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF
Options: R&S ZNA-K28
Manual operation: See "Controlled Timing" on page 566
SCPI command reference