Command reference
R&S
®
ZNA
1367User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure <CalType>,
<Characterization>[, <CalName>]
Selects a calibration type and a cal unit characterization (cal kit file) for an automatic
calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalType> FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | PFNPort | VUOSm | VPUosm |
CPUosm | UMDelay | PUMDelay
Calibration Type
FNPort
Full n-port (UOSM) calibration
TOSM
Full n-port calibration with characterized Through.
If the selected cal unit characterization does not contain a
Through characterization, the command silently falls back to a
FNPort (UOSM) calibration.
FOPort
Full one-port ("Refl OSM") calibration
OPTPort
"One Path Two Ports" calibration
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
FRTRans
Transmission normalization, bidirectional
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
FTRans | RTRans
Transmission normalization, forward | reverse
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
REFL
Reflection normalization, Open
RSHort
Reflection normalization, Short
PFNPort
Full n-port (FNPort) calibration with additional power flatness
calibration
VUOSm
FNPort (UOSM) Vector Mixer calibration
VPUOSm
PFNPort (PUOSM) Vector Mixer calibration
CPUosm
PFNPort (PUOSM) Scalar Vector Mixer calibration