Command reference
R&S
®
ZNA
1368User Manual 1178.6462.02 ─ 29
UMDelay
UOSM + Delay Mixer calibration for two-tone group delay mea-
surements
PUMDelay
UMDelay with complementary source flatness calibration
<Characterization> Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
– The empty string (' ') refers to the factory calibration of (and
stored on) the active calibration unit.
– A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
– A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
– A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
Note that for inline calibration systems this setting is ignored.
The suitable, ICU-specific characterizations must be specified
using SYSTem:COMMunicate:RDEVice:AKAL:CKIT.
Setting parameters:
<CalName>
Example:
SENSe1:CORRection:COLLect:AUTO:CONFigure
FNPort,''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
Example: See Chapter 8.2.5.6, "Vector mixer calibration using calibration
units", on page 1754
Manual operation: See "Calibration Type" on page 585
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt <PortNumber>
Defines the "Common Port" of an automatic calibration.
This port will be part of every port assignment that is defined using [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault, or [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
on page 1363.
If [SENSe:]CORRection:COLLect:AUTO:MCONnect is ON, this port will be used as
the center of all Through measurements. If you set up the port assignments manually,
make sure that every port assignment contains it.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<PortNumber> Number of the common port
SCPI command reference