Command reference
R&S
®
ZNA
1362User Manual 1178.6462.02 ─ 29
Test Port Port Assignment 1 Port Assignment 2
2 Cal Unit Port 1 -
3 Cal Unit Port 2 auto-detected
4 - auto-detected
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
<TestPort1>[, <TestPort2>]...
Creates the default port assignment(s) for the specified test ports.
The number of created assignments can be queried using the [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:COUNt? command.
Note that during the corresponding calibration sweep the R&S ZNA expects the physi-
cal port connections to be established exactly as specified by the created port assign-
ments.
Use
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:
DEFault on page 1363 to use auto-detection instead of defining "test port --> cal
unit port"-connections explicitely
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine or
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt to define (additional) port assignments manually
See Chapter 4.5.5.6, "Multiple port assignments", on page 216 for background infor-
mation.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TestPort1> Test port number 1
<TestPort2> Test port number 2
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1360
Usage: Setting only
Manual operation: See "CalUnit Con" on page 588
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
<TestPort1>[, <TestPort2>]...
Similar logic as [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
DEFine, but initially defines the port assignment only by its underlying test port set; the
connected calibration unit ports are auto-detected at the start of the corresponding
calibration sweep.