Command reference
R&S
®
ZNA
1369User Manual 1178.6462.02 ─ 29
Usage: Setting only
Manual operation: See "Common Port" on page 585
[SENSe:]CORRection:COLLect:AUTO:MCONnect <MinimumConnect>
Enables or disables the Reduced Through logic for every port assignment of an auto-
matic calibration.
Note that for this to work, the port assignments must have a star topology, with a
"Common Port" set using [SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
on page 1368.
Setting parameters:
<MinimumConnect> ON (1), or OFF (0)
Manual operation: See "Use Reduced Number of Through" on page 859
[SENSe<Ch>:]CORRection:COLLect:AUTO:MDELay <Port1>, <Port2>
Performs the delay mixer calibration step as part of an automatic calibration
Suffix:
<Ch>
.
Number of the calibrated channel
Setting parameters:
<Port1> RF port
<Port2> IF port
Usage: Setting only
Options: R&S ZNA-K9
[SENSe<Ch>:]CORRection:COLLect:AUTO:MIXer <Port1>, <Port2>
Acquire calibration data for the calibration mixer in an automatic vector mixer calibra-
tion
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Port1> VNA port connected to the RF port of the calibration mixer
<Port2> VNA port connected to the IF port of the calibration mixer
Example: See Chapter 8.2.5.6, "Vector mixer calibration using calibration
units", on page 1754
Usage: Setting only
Options: R&S ZNA-K5
SCPI command reference