Command reference
R&S
®
ZNA
1609User Manual 1178.6462.02 ─ 29
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZV-Z52::1234'
Select the cal unit named 'ZV-Z52::1234' for calibration.
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port TOSM calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Manual operation: See "Calibration Unit" on page 622
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
Queries the names (USB addresses) of all connected calibration units.
Example:
See SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Usage: Query only
Manual operation: See "Calibration Unit" on page 622
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:SUBModule:ALL?
If the calibration unit selected using SYSTem:COMMunicate:RDEVice:AKAL:
ADDRess is an inline calibration system (ICS), this query returns the addresses of the
connected ICUs. Otherwise it returns the empty string.
Example:
:SYSTem:COMMunicate:RDEVice:AKAL:
ADDRess 'ZN-Z30-02::100010'
Selects an ICS via its Inline Calibration Controller (ICC)
R&S ZN-Z30 (serial number 100010).
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:
SUBModule:ALL?
Returns something like:
'[1]ZN-Z33-02::101325,[2]ZN-Z33-02::
101325,[3]ZN-Z33-02::101339'
i.e. there are three ICUs of type R&S ZN-Z32 variant 02 connec-
ted to the ICC. For correct port assignment during calibration,
the ICU with serial number 101325 is considered port 1, 101325
port 2, and 101339 port 3 of the ICS.
Usage:
Query only
Manual operation: See "Configured Devices" on page 927
SYSTem:COMMunicate:RDEVice:AKAL:CKIT <Characterization>
If the calibration unit selected using SYSTem:COMMunicate:RDEVice:AKAL:
ADDRess is an inline calibration unit (ICU), this command sets/queries the characteri-
zation to be used for calibrations with this ICU.
SCPI command reference