EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1609

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1609User Manual 1178.6462.02 ─ 29
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZV-Z52::1234'
Select the cal unit named 'ZV-Z52::1234' for calibration.
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port TOSM calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Manual operation: See "Calibration Unit" on page 622
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
Queries the names (USB addresses) of all connected calibration units.
Example:
See SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Usage: Query only
Manual operation: See "Calibration Unit" on page 622
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:SUBModule:ALL?
If the calibration unit selected using SYSTem:COMMunicate:RDEVice:AKAL:
ADDRess is an inline calibration system (ICS), this query returns the addresses of the
connected ICUs. Otherwise it returns the empty string.
Example:
:SYSTem:COMMunicate:RDEVice:AKAL:
ADDRess 'ZN-Z30-02::100010'
Selects an ICS via its Inline Calibration Controller (ICC)
R&S ZN-Z30 (serial number 100010).
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:
SUBModule:ALL?
Returns something like:
'[1]ZN-Z33-02::101325,[2]ZN-Z33-02::
101325,[3]ZN-Z33-02::101339'
i.e. there are three ICUs of type R&S ZN-Z32 variant 02 connec-
ted to the ICC. For correct port assignment during calibration,
the ICU with serial number 101325 is considered port 1, 101325
port 2, and 101339 port 3 of the ICS.
Usage:
Query only
Manual operation: See "Configured Devices" on page 927