Command reference
R&S
®
ZNA
1608User Manual 1178.6462.02 ─ 29
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator <Terminator>
Sets the receive terminator of the analyzer. The receive terminator indicates the end of
a command or a data block.
The receive terminator setting is relevant if block data is transferred to the analyzer
(FORMat[:DATA] REAL). In the default setting LFEoi, the analyzer recognizes an LF
character sequence with or without the EOI control bus message as a receive termina-
tor. An accidental LF in a data block can be recognized as a terminator and cause an
interruption of the data transfer.
The EOI setting is especially important if commands are transferred in block data for-
mat, because it ensures that the parser for command decoding is activated by the ter-
minator only after the command has been completely transferred. Readout of binary
data does not require a change of the receive terminator.
Parameters:
<Terminator> LFEoi | EOI
LFEoi
A line feed character sequence with or without EOI is recognized
as receive terminator
EOI
Only EOI is recognized as receive terminator
Example:
SYST:COMM:GPIB:RTER EOI
Set the terminator to EOI.
SYSTem:COMMunicate:NET:HOSTname <HostName>
Sets or gets the host name of the instrument.
Parameters:
<HostName> Host name
*RST: A preset of the instrument doesn't change the host
name.
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess <Address>
Selects one of the USB-connected calibration units for calibration (see commands
SENSe<Ch>:CORRection:COLLect:AUTO...). This command is not necessary if
only one cal unit is connected.
Parameters:
<Address> Name (USB address) of a connected calibration unit (string vari-
able). The names of all connected cal units can be queried using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?.
SCPI command reference