Command reference
R&S
®
ZNA
1371User Manual 1178.6462.02 ─ 29
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Returns the assignment between the physical analyzer ports and the ports of the con-
nected automatic calibration unit.
Example:
CORR:COLL:AUTO:PORT '', 1, 2, 2, 4, 4, 1
Perform an automatic 3-port calibration at the analyzer ports 1,
2, and 4 using the calibration unit's default calibration kit file and
ports 2, 4, and 1 of the cal unit.
CORR:COLL:AUTO:PORT:CONN?
Query the actual port assignment. If the cal unit is properly con-
nected according to the previous command, the response is
1,2,2,4,3,0,4,1. A zero means that the corresponding analyzer
port is not connected to any port of the calibration unit.
Usage: Query only
Manual operation: See "Detect Port Assignment" on page 581
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
<CalType>[,
<Characterization>[, <TestPort1>[, <CalUnitPort1>[, <TestPort2>]...]]]
Selects and initiates an automatic calibration at arbitrary test and calibration unit ports.
A progress monitor for the calibration sweeps is displayed.
Tip: This command is necessary if the analyzer fails to detect the cal unit ports con-
nected to each of its ports (e.g. because of a high attenuation in the test setup). If auto-
detection works you can use the simpler command [SENSe<Ch>:]CORRection:
COLLect:AUTO:TYPE.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType> FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | PFNPort | VUOSm | CPUosm |
UMDelay | PUMDelay
See [SENSe<Ch>:]CORRection:COLLect:AUTO:
CONFigure on page 1367.
<Characterization> Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
If an empty string (' ') is specified, the factory cal kit file
stored in the active calibration unit is used.
A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.