EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 581

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
581User Manual 1178.6462.02 ─ 29
If switch matrices are involved, the label also indicates:
which VNA port b is switched to the input port j and
which VNA port a is switched to the output port i
Instead of a single Sij trace we have multiple traces Sijvativbtj. E.g., "S21v1t1v3t2"
means that S21 is measured with VNA port 1 connected to the input port and VNA port
3 connected to the output port.
Power Sweep Diagrams
Power sweep traces are displayed during the "Power" calibration sweeps of a SMAR-
TerCal.
Start Cal Sweep / Abort Sweep
Starts the calibration sweep for the related port assignment or aborts it.
Note: The power calibration sweep is performed at the "Reference Receiver Cal
Power" level specified in the Cal Power Config dialog.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
Detect Ports & Start Cal
Performs the "Detect Port Assignment" and "Start Cal Sweep" functions, one after the
other.
Detect Port Assignment
Starts a procedure by which the R&S ZNA (with a little help from the attached calibra-
tion unit) auto-detects the connected ports. The automatic assignment replaces the
configured one.
In case auto-detection fails
an error report is shown as a warning dialog
the undetected port connections are marked with warning signs
the calibration can be invalid
Cal softtool