Command reference
R&S
®
ZNA
1372User Manual 1178.6462.02 ─ 29
<TestPort1> Test port number. For an n-port automatic calibration, n arbitrary
(not necessarily consecutive) port numbers must be specified.
For a one path two port calibration (OPTPort), the first port no.
denotes the source port (fully calibrated port).
<CalUnitPort1> Port numbers of the cal unit. For an n-port automatic calibration,
n arbitrary (not necessarily consecutive) port numbers must be
specified. It is possible to combine arbitrary (not necessarily
matching) pairs of analyzer and cal unit ports.
<TestPort2> Second test port number
*RST: n/a
Example:
CORR:COLL:AUTO:PORT FNPort, '', 1, 2, 2, 4, 4,
1
Perform an automatic full 3-port calibration at test ports 1, 2, and
4 using the calibration unit's default calibration kit file and ports
2, 4, and 1 of the cal unit.
See also [SENSe<Ch>:]CORRection:COLLect:AUTO:
POWer
Usage: Setting only
Manual operation: See "Apply" on page 582
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer <TestPort>
Initiates a receiver power calibration at physical port <TestPort>. This is the second
step of an automatic SMARTerCal.
Note: The power calibration must be performed "after" the system error correction; see
example below.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<TestPort> Test port number. The port number must match the port number
selected via [SENSe:]CORRection:COLLect:PMETer:ID;
see example.
SCPI command reference