Command reference
R&S
®
ZNA
1373User Manual 1178.6462.02 ─ 29
Example: Suppose that a power meter no. 1 is configured and USB-con-
nected to the analyzer.
SENSe:CORRection:COLLect:PMETer:ID 1, 2
Select power meter no. 1, to be connected to Port 2.
--> Connect the calibration unit between ports 1 and 2 of the
analyzer,
SENSe:CORRection:COLLect:AUTO:PORTs:TYPE
PFNPort, '', 1, 1, 2, 2
Perform an automatic full 2-port (UOSM) calibration at the ana-
lyzer test ports 1 and 2 using the calibration unit's default cali-
bration kit file and default port assignments (ports 1 and 2 of the
cal unit).
--> Remove the cal unit and connect the power meter to Port 2
of the analyzer.
SENSe:CORRection:COLLect:AUTO:POWer 2
Perform an additional receiver power calibration at Port 2. This
completes the
Usage:
Setting only
Manual operation: See "Start Cal Sweep / Abort Sweep" on page 581
[SENSe<Ch>:]CORRection:COLLect:AUTO:REPeat
Use this command to reuse measurement data of the active automatic calibration (if
available) for the next automatic calibration. This data reuse allows an automatic cali-
bration to be repeated without repeating all measurements.
Reusing existing calibration data is particularly useful for inline calibrations, whose
OSM measurements can be repeated without reconnections, but whose unknown
through measurements can not.
Note that this command does not work with "single-command calibrations" performed
using [SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE.
Suffix:
<Ch>
.
Channel number
Example: See Chapter 8.2.5.5, "Inline calibration", on page 1754
Usage: Event
Manual operation: See "Repeat Cal" on page 589
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Calculates the system error correction for automatic calibrations with Multiple port
assignments, saves the data and applies the calibration to the active channel.
Requires successful calibration sweeps for all related port assignments (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire)
Suffix:
<Ch>
.
Channel number of the calibrated channel
SCPI command reference