Command reference
R&S
®
ZNA
1347User Manual 1178.6462.02 ─ 29
[SENSe:]CORRection:CKIT:SELect <ConnectorType>[, <CalKitName>]
Selects the calibration kit to be used, specifying its connector type and name
(optional).
Tip: The command is suitable for connector types with arbitrary, user-defined names.
For standard connector types you can use the command [SENSe:]CORRection:
CKIT:<ConnType>:SELect.
Parameters:
<ConnectorType> Connector type, e.g. a user-defined connector type (string vari-
able).
<CalKitName> String parameter containing the name of a calibration kit availa-
ble on the analyzer.
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\ZNA\
Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:SEL 'N 50 Ohm', 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in New Kit 1.calkit reads New
Kit 1.
Manual operation: See "Cal Kit Port <i>" on page 587
[SENSe:]CORRection:CKIT:<ConnType>:SELect <CalKitName>
Selects the calibration kit to be used for a specified connector type <ConnType>. The
kit is identified by its name.
Tip: For connector types with arbitrary, user-defined names you can use the command
[SENSe:]CORRection:CKIT:SELect.
Cal kit naming conventions
Calibration kit names must be entered as string parameters. The string contains the cal
kit name used in the calibration dialogs (e.g. "Calibration Presettings"); a "Ω" in the
name must be replaced by Ohm, e.g.:
●
NewKit1 denotes the user-defined calibration kit "NewKit1".
●
N 50 Ohm Ideal Kit denotes the "N 50 Ω Ideal Kit".
●
ZV-Z21 typical denotes the cal kit "ZV-Z21 typical".
Suffix:
<ConnType>
.
Connector type, one of the following identifiers:
N50, N75 – N 50 Ω or N 75 Ω connectors
PC7, PC35, PC292 – PC 7, PC 3.5 or 2.92 mm connectors
USER<no> – user-defined connectors UserConn1, UserConn2
SMA – user-defined connector type SMA ...
SCPI command reference