Command reference
R&S
®
ZNA
1348User Manual 1178.6462.02 ─ 29
Parameters:
<CalKitName> String parameter containing the name of a calibration kit availa-
ble on the analyzer. See "Cal kit naming conventions " above.
*RST: n/a (A *RST does not change the assignment
between connector types and calibration kits.)
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\ZNA\
Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:N50:SEL 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in New Kit 1.calkit reads New
Kit 1.
Manual operation: See "Add / Copy / Delete / Standards..." on page 617
[SENSe:]CORRection:CKIT:STANdard:CATalog? <CalKitName>
Returns a list of all standards in a given calibration kit.
Query parameters:
<CalKitName> Name of the cal kit. Use [SENSe:]CORRection:CKIT:
CATalog? to query cal kit names.
Example: See [SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage: Query only
[SENSe:]CORRection:CKIT:STANdard:DATA? <ConnectorType>, <CalKitName>,
<KitLabel>, <StandardType>, <DelayMode>[, <Port1>[, <Port2>]]
Returns the data of the related calibration standard (identified by <ConnectorType>,
<CalKitName>, <KitLabel> and <StandardType>) and - optionally - restricted to the
given port(s).
The delay parameter is returned according to the selected <DelayMode>; see
Table 7-14.
Query parameters:
<ConnectorType> Name of the connector type. Use [SENSe<Ch>:
]CORRection:CONNection:CATalog? to query connector
names.
<CalKitName> String parameter containing the name of the calibration kit. See
"Cal kit naming conventions " under [SENSe:]CORRection:
CKIT:<ConnType>:LSELect.
<KitLabel> String parameter containing the label of the calibration kit, usu-
ally the serial number.
SCPI command reference