1345User Manual 1178.6462.02 ─ 29
<std_type> Meaning
MSMatch | FSMatch Sliding match: m or f
MREFlect | FREFlect Reflect: m or f
MMTHrough | MFTHrough |
FFTHrough
Through: m-m or m-f or f-f
MMLine[<1|2|3>] | MFLine[<1|2|3>]
| FFLine[<1|2|3>]
MMLine[<1|2|3>](P2P3) ...
Line: m-m or m-f or f-f (three standards each)
Suffix 1 can be omitted.
MMATten | MFATten | FFATten Attenuation: m-m or m-f or f-f
MMSNetwork | MFSNetwork |
FFSNetwork
Symmetric network: m-m or m-f or f-f
[SENSe:]CORRection:CKIT:CATalog? [<ConnectorType>]
Returns a list of all cal kits for a given connector type or for all connector types.
Query parameters:
<ConnectorType> Name of the connector type (optional). Use [SENSe<Ch>:
]CORRection:CONNection:CATalog? to query connector
names.
If omitted, the command returns the list of all cal kits.
Example: See [SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage: Query only
Manual operation: See "Available Cal Kits" on page 616
[SENSe:]CORRection:CKIT:DELete <CalKitName>
Deletes an imported or user-defined cal kit.
Note: It is not possible to modify or store predefined or ideal kits.
Setting parameters:
<CalKitName> String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
Example: See [SENSe:]CORRection:CKIT:LABel
Usage: Setting only
Manual operation: See "Add / Copy / Delete / Standards..." on page 617