Command reference
R&S
®
ZNA
1343User Manual 1178.6462.02 ─ 29
Table 7-13: Set command parameters
One port standards Two port standards
Open (Offset)
Short
Match Reflect Sliding
match
Symmetric
network
Through,
Line
Attenuation
<Con-
nType>
mandatory
<CalKit-
Name>
mandatory (and must not be an empty string)
<Standar-
dLabel>
mandatory (but can be an empty string)
<MinFreq>,
<MaxFreq>
mandatory
<Delay-
Param>,
<Loss>,
<Z0>
mandatory mandatory mandatory mandatory not used mandatory not used
<C0>,
<C1>,
<C2>, <C3>
mandatory mandatory* mandatory* mandatory** mandatory** not used
<L0>, <L1>,
<L2>, <L3>
mandatory* mandatory mandatory* mandatory** mandatory**
OPEN |
SHORt |
MATCh |
<Resist-
ance>
optional*** optional*** optional*** mandatory:
OPEN |
SHORt |
<Resist-
ance>
mandatory:
OPEN |
SHORt |
<Resist-
ance>
<Port1>[,
<Port2>]
optional: <Port 1> optional: <Port 1>,<Port 2>
* values are ignored during calibration
** if OPEN is selected the residual inductance (L0,...,L3) are during calibration; if SHORt is selected the fringing capacitance
(C0,...,C3) are during calibration
*** must be provided if the definition is restricted to a single port. Values are ignored during calibration
The parameters in the [SENSe<Ch>:]CORRection:CKIT:<StandardType>,
[SENSe<Ch>:]CORRection:CKIT:<StandardType>:WLABels, and
[SENSe<Ch>:]CORRection:CKIT:<ConnectorType>:<StandardType> com-
mands have the following meaning:
Table 7-14: Parameter list
Parameter Meaning Comment/Unit
'<ConnType>' Name of the connector type.
Use SENSe1:CORRection:CONNection:CATalog? to query connec-
tor names.
String parameter
'<CalKitName>' Name of the calibration kit. Use
[SENSe:]CORRection:CKIT:CATalog? to query cal kit names.
String parameter