Command reference
R&S
®
ZNA
1394User Manual 1178.6462.02 ─ 29
Usage: Event
Manual operation: See "Apply" on page 578
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]
Calculates the system error correction data from the acquired measurement results
([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected), stores it and
applies it to the calibrated channel(s). To avoid incompatibilities, older system error cor-
rection data is deleted unless it has been transferred into a cal pool (MMEMory:
STORe:CORRection <Ch>, '<file_name>').
Suffix:
<Ch>
.
One of the calibrated channels
If [SENSe:]CORRection:COLLect:CHANnels:ALL is ON
this suffix can be omitted
Example:
CORRection:COLLect:METHod:DEFine 'Test',RSHort,
1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORRection:COLLect:CHANnels:ALL ON
Enable calibration in all channels of the active recall set.
CORRection:COLLect:SELected SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORRection:COLLect:SAVE:SELected
Calculate the system error correction data and apply it to all
channels of the active recall set.
Usage:
Event
Manual operation: See "Apply" on page 578
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt> <Type>[, <Gender>]
Selects a connector type at a specified port <PhyPt> and its gender. In contrast to
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>, this command uses
a string variable to identify the connector type.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt> Port numbers of the analyzer. If the analyzer is set to use the
same connectors at all ports ([SENSe<Ch>:]CORRection:
COLLect:CONNection:PORTs ALL), then a change of a con-
nector type is valid for all ports. The gender of the connectors
can still be different.
SCPI command reference