Command reference
R&S
®
ZNA
1356User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected.................................................... 1387
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine....................................................1388
[SENSe<Ch>:]CORRection:COLLect:NFIGure[:ACQuire].................................................1390
[SENSe<Ch>:]CORRection:COLLect:NFIGure:END........................................................ 1391
[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE.......................................................1391
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt......................................................1392
[SENSe:]CORRection:COLLect:PMETer:ID.....................................................................1392
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault........................................ 1393
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]...................................... 1394
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>............................................ 1394
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected <Type>, <TestPort>[,
<SecondPortOrAdapter>[, <Dispersion>[, <DelayTimePhase>]]]
Starts the acquisition of measurement data for the selected standard and port(s). The
standards are reflection or transmission standards and can be connected to arbitrary
analyzer ports.
Note:
●
The calibration measurement has a variable timeout: Timeout = (Sweep time /
Number of sweep points) * 3 + 0.1 s
The timeout may be reached e.g. while the measurement waits for the specified
trigger event. It is used for all standard measurements including power calibration
sweeps.
●
For a sliding match, the R&S ZNA can acquire measurement data for up to 20
positions per port. Multiple calls of
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected SLIDe,
<TestPort> for the same test port, implicitly increases the position until position
20 has been recorded. Subsequent calls will start over at position 1, overwriting the
previously acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<Type> THRough | OPEN | SHORt | OSHort | OSHort1 | OSHort2 |
OSHort3 | MATCh | NET | ATT | REFL | SLIDe | ISOLation |
LINE | LINe1 | LINe2 | LINe3 | UTHRough | POWer | MIXer |
TERMination | EATTenuator | MDELay
Standard types: Through, Open, Short, Match, Symmetric Net-
work (NET), Attenuation (ATT), Reflect, Sliding Match (SLIDe),
Line1 (LINE1 and LINE are synonymous), Line2 and Line3 (esp.
for TRL calibration), Offset Short 1 to 3 (OSHort), Unknown
Through, power meter (for SMARTerCal), Mixer.
ISOLation is not a physical standard: to measure the isolation
(supported for transmission normalization and TOSM only!) it is
recommended to suitably terminate the related test ports (e.g.
with 50 Ω loads).
MIXer represents the calibration mixer measurement that is
required for calibrating a mixer phase measurement (with option
R&S ZNA-K5).
SCPI command reference