Command reference
R&S
®
ZNA
1387User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Channel number of the corrected channel
Example: See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
Usage: Event
Manual operation: See "Measure Fixture wizard" on page 720
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected <CalGroupFile>,
<Standard>, <TestPort1>[, <SecondPortOrAdapter>]
Reloads a set of previously acquired calibration data for a particular standard from a
file in the cal pool. The loaded data may be combined with new calibration measure-
ment data ([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected) in order
to simplify and speed up the new calibration procedure. The channel settings for loa-
ded and new calibrations (e.g. the number of sweep points) must be identical.
Note:
●
The analyzer performs a consistency check for the loaded data. If the loaded file is
incompatible with the channel settings of channel <Ch>, or if it does not contain
data for the specified standard and port(s), a command error message (–100,
"Command error;...") is generated.
●
For a sliding match, the R&S ZNA can acquire and load measurement data for up
to 20 positions per port. Multiple calls of
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected <CalGroupFile>,
SLIDe, <TestPort> for the same cal group file and test port, implicitly increases
the position until position 20 has been recorded. Subsequent calls will start over at
position 1, overwriting the previously loaded data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalGroupFile> String parameter to specify the name of the loaded cal group
file.
<Standard> THRough | OPEN | SHORt | OSHort | OSHort1 | OSHort2 |
OSHort3 | MATCh | NET | ATT | REFL | SLIDe | ISOLation |
LINE | LINe1 | LINe2 | LINe3 | UTHRough | POWer | MIXer |
TERMination | EATTenuator | MDELay
See [SENSe<Ch>:]CORRection:COLLect[:ACQuire]:
SELected on page 1356.
<TestPort1> Test port number. For a transmission standard (through, line,
attenuation, symmetric network) or an adapter used as a
"through" the input and output port numbers must be specified.
For reflection standards, only one port number is required.
SCPI command reference