Command reference
R&S
®
ZNA
1383User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]CORRection:COLLect:DETector <CalName>[, <CalDetectorType>]
Selects the detector type during manual calibration <CalName>.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName> Name of a manual calibration previously created e.g. using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalDetectorType> NORMal | AVERage
AVERage
Average detector, recommended for channels measuring traces
using detectors other than the NORMal one
Automatically selected for noise figure calibrations (see
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt).
NORMal
Normal (sample) detector, recommended for all other applica-
tions
*RST: NORMal
Manual operation: See "Detector" on page 645
[SENSe<Ch>:]CORRection:COLLect:DISCard
Terminates a system error correction, discarding the acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
:SENSe1:CORRection:COLLect:METHod:DEFine 'Test', TOSM, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected THRough, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected SHORt, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 2
:SENSe1:CORRection:COLLect:DISCard
Usage: Event
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire] <StandardType>,
<TestPort1>[, <TestPort2>]...
Starts a fixture compensation sweep in order to acquire measurement data for a test
fixture that has its inner conductor terminated with the selected standards.
Suffix:
<Ch>
.
Channel number of the corrected channel
SCPI command reference