GUI reference
R&S
®
ZNA
645User Manual 1178.6462.02 ─ 29
Only calibration types supported by METAS VNA Tools are available.
Ports
Selects the test ports to be calibrated.
Note: Calibration and port de-/activation.
The analyzer fimware automatically activates/deactivates ports during/after a (success-
ful) calibration:
Calibrated ports that were previously disabled, are automatically enabled as single-
ended logical ports.
An uncalibrated port that is not used by a measurement (i.e. the port is not required
by any trace of the related channel) is disabled.
Remote command:
The port parameters in many calibration commands define the calibrated ports.
Type
Selects the calibration type.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine <CalName>, MTOSm |
MUOSm | MOSM | MTRL, ...
Detector
Selects the detector type the analyzer uses during the calibration.
"AVG"
The average detector eliminates noise contributions which are super-
imposed on the measured signal.
If an S-parameter, wave quantity or ratio is measured with a "Detec-
tor" on page 350 other than "Normal", AVG is selected by default.
The "AVG" detector is used for Noise figure measurement.
The "Normal" detector is recommended for all other applications.
Cal softtool