Command reference
R&S
®
ZNA
1382User Manual 1178.6462.02 ─ 29
[SENSe:]CORRection:COLLect:CSETup <Boolean>
Defines how calibration sweeps are prepared and performed.
This setting is valid for manual and automatic calibration.
See [SENSe<Ch>:]CORRection:COLLect:CHANnels:CSETup for a channel-spe-
cific variant of this setting.
Parameters:
<Boolean> OFF (0)
Individual sweep setups for all standards (faster for some config-
urations)
ON (1)
Common sweep setup for all standards (no preparation phase
for each standard)
Note that noise figure calibrations require individual setups, and
hence this setting is ignored.
*RST: OFF
Manual operation: See "Same Sweep Setup for All Standards" on page 859
[SENSe<Ch>:]CORRection:COLLect:DELete [<CalName>]
Deletes system error correction data generated and stored previously.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalName> Name of the calibration (string parameter) defined together with
the calibration type ([SENSe<Ch>:]CORRection:COLLect:
METHod:DEFine).
ALL - the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected[:DUMMy].
*RST: -
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
CORR:COLL:DEL ALL
Calculate the system error correction data and apply it to the
active channel, then delete the data.
Usage: Setting only
Manual operation: See "Apply" on page 578
SCPI command reference