Command reference
R&S
®
ZNA
1417User Manual 1178.6462.02 ─ 29
Increasing (decreasing) the values in the correction data sets increases (decreases)
the input power reading. Writing correction data (the setting command) fails if the num-
ber of transferred values is not equal to the number of sweep points.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S ZNA-K4).
Parameters:
<Wave> Identifier for the calibrated wave:
'A<n>' or 'AWAV<n>' denote correction data for the reference
wave a<n>, where <n> corresponds to the port number.
'B<n>' or 'BWAV<n>' denote correction data for the reference
wave b<n>.
<CorrData> Power correction values either in ASCII or block data format,
depending on the current FORMat[:DATA] setting.
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:DATA? 'AWAV1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
CORR:POW:DATA 'AWAV1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.
Manual operation:
See "Start Cal Sweep" on page 602
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire
Starts the receiver calibration
●
at physical port <PhyPt>
●
in channel <Ch>
●
at harmonic orders [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?
?
Then stores and applies the calibration data.
Note: The receiver calibration relies on the source power calibration acquired using
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire].
Suffix:
<Ch>
.
Channel number
<PhyPt> Port number.
SCPI command reference