Command reference
R&S
®
ZNA
1338User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?..............................................1338
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD........................................1338
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove.................................. 1339
[SENSe<Ch>:]CORRection:ADVanced:POWer<PhyPt>...................................................1339
[SENSe<Ch>:]CORRection:ADVanced:POWer<PhyPt>:STATe.........................................1339
[SENSe<Ch>:]CORRection:ADVanced:SENSe<PhyPt>:ATTenautor................................. 1339
[SENSe<Ch>:]CORRection:ADVanced:SENSe<PhyPt>:ATTenautor:STATe....................... 1340
[SENSe<Ch>:]CORRection:ADVanced:SENSe<PhyPt>:PREamp.....................................1340
[SENSe<Ch>:]CORRection:ADVanced:SENSe<PhyPt>:PREamp:STATe...........................1340
[SENSe<Ch>:]CORRection:ADVanced:SOURce<PhyPt>:ATTenuator............................... 1340
[SENSe<Ch>:]CORRection:ADVanced:SOURce<PhyPt>:ATTenuator:STATe..................... 1341
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?
Returns the list of harmonic orders at which the source and receiver power calibration
shall be performed for harmonics measurements in channel <Ch>.
Use [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD/
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove to add/remove
harmonic orders to/from this list.
Use SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire] and
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire) to calibrate.
Suffix:
<Ch>
.
Channel number
Usage: Query only
Options: R&S ZNA-K4
Manual operation: See "Harmonic Orders to Calibrate" on page 590
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD <Order>
Adds a harmonic order to the list of harmonic orders at which source and receiver cali-
brations shall be performed for harmonics measurements in channel <Ch>.
Use [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove to get this
list, and [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove to
remove a harmonic order from this list.
Use SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire] and
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire) to calibrate.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Order> Order number
Usage: Setting only
Options: R&S ZNA-K4
SCPI command reference

Table of Contents

Related product manuals