Command reference
R&S
®
ZNA
1414User Manual 1178.6462.02 ─ 29
For query:
1 - at least one length offset parameter is different from its
default value
0 - all length offsets are zero / set to default
*RST: OFF
Example:
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset param-
eters have been reset as well. The response is 0.
Manual operation: See "Reset Offsets" on page 708
[SENSe<Ch>:]CORRection:PCAL <ActivatePowerCals>
Activates or deactivates all power calibrations.
Suffix:
<Ch>
.
Calibrated channel number
Setting parameters:
<ActivatePowerCals> NONE | ALL
NONE deactivate all power calibrations.
ALL activate all power calibrations.
*RST: n/a
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is 1).
CORR:PCAL NONE; POW:AWAV?
Disable all power calibrations and check again whether the cali-
bration is applied (the response is 0).
Usage: Setting only
Manual operation: See "Scalar Power Cal – All Power Cals On / All Power Cals
Off" on page 636
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire <Wave>[, <CalPort>[,
<SourceType>[, <SourcePort>[, <CalOnlyPortFreq>]]]]
Selects the wave quantity and the source for the receiver power calibration, starts the
calibration sweep, and applies the receiver power correction.
Suffix:
<Ch>
.
Calibrated channel number
SCPI command reference