Command reference
R&S
®
ZNA
1415User Manual 1178.6462.02 ─ 29
<PhyPt> This suffix is ignored because the port number is specified in the
parameter list.
Setting parameters:
<Wave> AWAVe | BWAVe | B1 | B2 | B3 | B4
AWAVe – calibration of reference waves a
1
, a
2
, .... In manual
control, the reference receiver calibration is included in the
source power calibration of each port.
BWAVe – calibration of received wave b
1
, b
2
, .... The port num-
ber <CalPort>, the used <SourceType>, and the
<SourcePort> number must be specified in addition.
Alternative: Parameters B1 and B2.
B1 | B2 ... – direct wave and cal port setting for received waves
b
1
, b
2
... The parameters B1 and B2 are ZVR-compatible. No
additional parameters must be specified. The source for B1 is
port 2 and vice versa.
<CalPort>
Calibrated port number
Range: 1 to port number of the analyzer
<SourceType> PORT | GENerator
PORT - internal source at port <CalPort>, to be fed to port
<SourcePort> using an external through connection. If
<CalPort> = <SourcePort>, an Open or Short standard is
required.
<SourcePort> Number of the port for the internal source or generator.
<CalOnlyPortFreq> Relevant for frequency conversion measurements only.
OFF | 0
The reference receiver at port <CalPort> is calibrated for all fre-
quencies that are relevant for any of the ports.
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST: 0
SCPI command reference